4 Mounting
Standpipes
1
Fig. 9: Mounting in a bypass tube - Position of the spacer or the centering weight
1
Rod probe with spacer (PEEK)
2
Cable probe with centering weight
3
Spacer (PEEK) on the gravity weight of a cable probe
h
Measurable tube section
Note:
Measurement in a standpipe is not recommended for extremely adhe-
sive products. In case of slight buildup, you should choose a bypass
tube with bigger diameter.
Instructions for the measurement:
connection to the vessel.
connection to the vessel.
A false signal suppression with installed sensor is generally rec-
ommended to achieve the best possible accuracy.
Standpipes or surge pipes are normally metal tubes with a diameter
in), such a pipe corresponds to a coax measuring probe. It does not
matter if the standpipe is perforated or slotted for better mixing.
Measuring probes can be mounted in standpipes up to DN 200.