7 Set up with the display and adjustment module
Measure the output current of BMD 1L with a suitable multimeter
and compare the measured output current with the calculated output
current.
You have to determine the permissible deviation of the values
yourself. This deviation depends on the the accuracy requirements of
your measurement loop. Determine the permissible tolerance for the
deviation.
If you have to interrupt the function, you can leave the BMD 1L in the
respective situation.
As long as BMD 1L is powered, the display and adjustment module
remains in the currently set adjustment menu.
To interrupt the function test, you have to push the button "ESC".
If you carry out the function test by means of the "PACTware" soft-
ware, you can store the previously performed tests and continue from
there later on.
Verify parameter
If the described process of parameter adjustment was run through
completely and correctly, the instrument will be locked and hence
ready for operation.
Otherwise the instrument remains in the released and hence unsafe
condition.
If you have to interrupt the function test, you can leave the display and
adjustment module of BMD 1L in its current state.
As long as BMD 1L is powered, the display and adjustment module
remains in the currently set adjustment menu.
To interrupt the function test, you have to push the button "ESC".
If you carry out the function test by means of the "PACTware" soft-
ware, you can store the previously performed tests and continue from
there later on.
Setup - Current output 2
If a supplementary electronics with an additional current output is in-
stalled in the instrument, you can adjust the additional current output
separately.
In menu item"Current output 2" you specify which measured value the
additional current output refers to.