Technical Data
12.3 Installation Conditions
12.3.1
Installation Site
For indoor use only, max. height above sea level
Temperature
No potentially explosive atmospheres
Access to operation-relevant parts is guaranteed
In operation
To ensure the metrological data
Relative humidity**
At temperatures up to 31°C, non-condensing
At temperatures of 31°C to 40°C, decreasing linearly
No heat from heating systems or direct sunlight
No electromagnetic fields
Storage, dry
* For conformity-assessed (verified) balances in accordance with EU requirements, refer to the information on the
balance.
** For conformity-assessed (verified) balances in accordance with EU requirements, the legal regulations apply.
12.3.2 Ambient Conditions at the Installation Site
Temperature
In operation
In operation, with isoCAL function, scope of application as per
Directive 2014/31/EU
In operation, without isoCAL function, scope of application as per
Directive 2014/31/EU
In operation, with conformity-assessed devices, as per specifications
on the device ID label
Relative humidity, in operation
At temperatures up to 31°C, maximum
Decreasing linearly thereafter, maximum
In operation, with conformity-assessed devices, as per specifications
on the device ID label
26
Operating Instructions Quintix
Pro
®
Unit
Value
m
3000
°C
+5 – +40
°C
+10 – +30
%
80
%
> 50
Unit
Value
°C
+5 – +40
°C
+10 – +30
°C
-17 – +17
%
80
%
> 50