Technical Data
12.11 Metrological Data
12.11.1 Models QTX324Ix | QTX224Ix | QTX124Ix
Model
Scale interval (d)
Maximum capacity (Max)
Weighing system
Repeatability
At 5% load, typical value
At approx. the maximum load, typi-
cal value
Linearity deviation
Limits
Typical value
Eccentricity deviation: Deviation when load is off-center, positions according to OIML R76
Test weight
Tolerance
Typical value
Sensitivity drift between +10°C and
+30°C
Tare maximum capacity (subtractive)
isoCAL:
Temperature change
Time interval
Only for models with approval:
Accuracy class
Type
Verification scale interval (e)
Minimum initial weight according to USP (United States Pharmacopeia), Chap. 41
Optimum minimum initial weight
Typical minimum initial weight
Typical measurement time
Typical stabilization time
30
Operating Instructions Quintix
Unit
Value
mg
0.1
g
320
EMC
± mg
0.08
± mg
0.1
± mg
0.2
± mg
0.06
g
200
± mg
0.4
± mg
0.2
± ppm/K
1
K
1.5
h
4
I
BC-QA
mg
1
g
0.082
g
0.16
s
≤2.0
s
≤1.5
Pro
®
Value
Value
0.1
0.1
220
120
EMC
EMC
0.08
0.08
0.1
0.1
0.2
0.2
0.06
0.06
100
50
0.4
0.4
0.12
0.1
1
1
1.5
1.5
4
4
I
I
BC-QB
BC-QB
1
1
0.082
0.082
0.16
0.16
≤2.0
≤2.0
≤1.5
≤1.5