What Should I Do If - Sirona SIROEndo Pocket Instrucciones De Uso

Ocultar thumbs Ver también para SIROEndo Pocket:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Operating Instructions SIROEndo Pocket
10
...the unit doesn't switch on
...the foot switch doesn't switch the
micromotor on
...the unit or the display is not
functioning properly
...the micromotor doesn't start
...the micromotor loses power
...the storage battery is not functioning
properly
... Calibration error 1 appears on the
display
61 97 722 D 3526
D 3526.201.01.05.02 07.2012

"What should I do if..."

Check whether the power supply connectors are plugged in properly.
1. Check whether the cable was plugged in correctly.
2. Check the motor operation by pressing the "Motor" key on the control
panel.
3. If the problem is not remedied by steps 1 and 2, the micromotor is defective
and must be replaced.
Execute a RESET: Press the "System " key while switching on the
SIROEndo Pocket or press and hold it for 1 second during the welcome
screen display.
CAUTION: If the unit switches off while the micromotor is running, this means
that the central unit is supplying the full amount of available power.
Checking the contra-angle handpiece.
Perform the following test steps:
1. Make sure that the micromotor connector is plugged into the unit correctly.
2. Check the contra-angle handpiece for damage and for correct fastening
to the micromotor.
3. Remove the contra-angle handpiece, switch to maximum speed and restart
the micromotor.
4. Remove the contra-angle handpiece and perform calibration.
5. Connect the contra-angle handpiece and perform calibration.
6. If the problem is not remedied by the preceding test steps, the micromotor
is defective and must be replaced.
1. Check the contra-angle handpiece for damage.
2. Perform calibration.
Check the following:
1. Whether the storage battery charges too quickly even though all measures
were followed correctly.
2. Whether the unit functions only when the power supply is connected to the
line power supply, and not in the battery-powered mode.
3. Whether the storage battery is defective. Have it replaced if necessary.
========================
Calibration error
Calibration error
1
========================
Check to make sure that the micromotor is properly connected to the control
unit.
10 "What should I do if..."
35

Publicidad

Tabla de contenido
loading

Tabla de contenido