11 Electromagnetic compatibility
11
Emission measurement
RF emissions according to
CISPR 11
RF emissions according to
CISPR 11
Harmonics
according to IEC 61000-3-2
Voltage fluctuations/Flicker
according to IEC 61000-3-3
38
Electromagnetic compatibility
11.1
Electromagnetic emission
The UNIT is intended for operation in the electromagnetic environment spec-
ified below.
The customer or user of the UNIT should make sure that it is used in such an
environment.
Conformity
Electromagnetic environment guidelines
Group 1
The UNIT uses RF energy only for its internal function. There-
fore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Class B
The UNIT is intended for use in all facilities, including residen-
tial areas and in any facilities connected directly to a public
power supply providing electricity to buildings used for resi-
not applicable
dential purposes.
Power output
< 50 W
not applicable
no significant
flicker