Oil-change
Over-used lubrication oil damages the pumps' mechanism; make sure to replace oil as soon as
it starts to turn cloudy. For conducting the oil-change, the pump must be switched off. For
changing the oil proceed as follows:
Unscrew the oil drainage plug located on the lower part of the pump
Check that oil comes out completely from the pump
Tightly screw the oil drainage plug back on
Re-fill the pump with oil (see "Filling of oil")
WARNING: Do not pollute the environment by spilling oil; the lubricant oil is a special
waste and must be disposed of according to relevant local law / guidelines in force.
8
Extraordinary maintenance / repair
Conduct extraordinary maintenance or repair whenever serious faults occur such as overheat-
ing, poor ultimate vacuum, persistent loudness / noise, blocked pump or any other irregular func-
tioning.
In that case, the pump needs to be disassembled, parts need to be cleaned, repaired or re-
placed. Only qualified service staff shall carry out such maintenance or repair work; otherwise
the warranty will be void.
For any information please contact our Service-After-Sales.
9
Troubleshooting
Problem
The pump doesn't
work
Overheating
Poor ultimate
vacuum
Persistent loud-
ness / noise
Blocked pump
Solution
Check that the pump is connected to the power supply.
Check that the plug and cable are not damaged.
Check that the power supply is suitable for the pump requirements (see
the label of the pump).
Check that the pump is filled up with oil (see section 2.1).
Contact our Service-After-Sales.
Check that the pump is filled up with oil (see section 2.1).
Check that the pump is not covered and that air could flow easily.
Check that the power supply is suitable for the pump requirements (see
the label of the pump).
Contact our Service-After-Sales.
Check that the pump has been working for the enough time. The final
vacuum level is directly related with the time of performance. The time
must be according to the volume to be evacuated.
Check that there is no leak in the connection or in the system to be
evacuated. Switch off the pump and check that the vacuum level re-
mains.
Check that all the connections and the equipment used are the appro-
priate ones.
Contact our Service-After-Sales.
Check that there are no damages on the fan cover. Any impact on this
area could affect the fan movement.
Check that the pump is perfectly srewed.
Contact our Service-After-Sales.
Contact our Service-After-Sales.