Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 25
3.
Support the instrument and connect the cables and pipes.
4.
Refer to the electrical connections below before fitting the instrument into the equipment
rack.
5.
Temporarily locate the two spigots* to each side of the equipment rack.
6.
Locate and slide the instrument into the rack.
7.
Locate the instrument on the spigots*.
8.
Secure the instrument in the equipment rack with two of the screws and washers (supplied).
9.
Remove the two spigots* and replace with the remaining two screws and washers (supplied).
3.2 Panel-mount Option
Refer to Figure C2 and the procedure below:
1.
Remove the four screws (1) from the instrument.
2.
Locate the instrument in panel mount assembly.
3.
Secure with the four screws (1).
4.
Support the instrument and connect the cables and pipes.
5.
Refer to the electrical connections below before fitting the instrument into the panel.
6.
Secure the instrument in the panel with four screws and washers (2).
4. Electrical Connections
RISK OF ELECTRIC SHOCK
electrical connections to the rear panel.
CAUTION
number IS1000118M9922-12). Using other power adaptors may cause over-
heating, this can result in a fire.
Do not let the power adaptor come into contact with any moisture or liquids.
4.1 Power Adaptor
1.
Before use, make sure the SELV power adaptor supplied with the instrument is used (Druck
part number IS1000118M9922-12). Refer to Figure A1, item 7.
2.
Install an accessible power isolator to use as the disconnecting device in the power adaptor
supply circuit.
3.
The power adaptor input power supply range: 100 - 240VAC, 50 to 60Hz, 600 - 300mA,
Installation Category II.
Note: The power adaptor must be supplied by a fused or overload-protected power supply.
4.
Connect the power adaptor to the instrument.
5.
Switch the power supply on.
6.
Check that the front panel display shows the power-up sequence (Refer to K0470,
Section 3.2, Power-up Sequence).
Note: After the power-up sequence, the instrument shows the default display on the touch
screen. The touch screen divides into a number of mimic keys.
4.2 Requirements for Rack-mounted and Panel-mounted Instruments
1.
Install an accessible power isolator to use as the disconnecting device in the power adaptor
supply circuit.
Use the power adaptor supplied with the instrument (Druck part
Isolate the power supply before making any
English–PACE1000 Safety Instructions | 7