If the cover needs replacing at any time, only use official
Swandoo Marie Integral fabrics.
DISINFECTION
Various products can be used for surface disinfection of
metal and plastic parts. Liquid disinfectants are available as
ready-to-use solutions that are sprayed on and evenly
applied with a soft cloth. Alternatively, wipes pre-soaked
with disinfectant can be used to wipe the products over the
entire surface. In both cases, care must be taken to ensure
complete wetting.
Disinfection in fully automatic disinfection systems is also
possible and recommended.
The exposure times may vary and can be found in the manu-
facturer's instructions for the products used.
SERVICING
Please carry out a visual inspection on a daily basis and
regularly check the product for cracks, breaks, missing parts
and malfunctions. In case of a defect or malfunction, please
contact the specialist dealer who supplied you with the
product.
MAINTENANCE
For reasons of user safety and to retain product liability, the
product must be subjected to maintenance by a specialist
dealer at least once a year. The maintenance work carried
out must be documented in the maintenance plan.
MAINTENANCE SPECIFICATIONS
Basic cleaning according to the manufacturer's specifications.
Disinfect according to the manufacturer's specifications as
required.
Damages to the frame, mounting parts and accessories
(cracks, breaks,corrosion, bent or missing parts).
Strength of the connections (tighten loose screws, replace
missing screws).
Functionality of the adjustment elements (screws, release lever).
Functionality of other adjustment elements (back, headrest,
pelotte pads, footrests and guides in the leg area).