Descargar Imprimir esta página

HT Instruments HT7051 Manual De Instrucciones página 52

Ocultar thumbs Ver también para HT7051:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

5.3.1 Configuration of the test parameters for ADJUST mode
a.
By cyclically pressing the SET/DISPLAY MEM key, it is possible to select the
following parameters:
• Minimum value of the insulation (MIN) in a range between 0,01M - MAX
• Maximum value of the insulation (MAX) in a range between (MIN) - 10T
• Value of the test voltage in a range between 100V - 5000V (can be set in steps of
25V up to 1000V and in steps of 50V from 1000V to 5000V)
• Test time value in a range between (00min:05sec) - (99min:59 sec).
b.
Press the arrow keys to change the value of the parameters (a long pressing of the
keys allows a rapid change of the values).
c.
Press the MODE/CLEAR SMOOTH key to remove the limit values set for the
parameters. The display shows the message "no" to indicate that there is no limit set.
Only when setting the value of the test voltage, by pressing the MODE/CLEAR
SMOOTH key, the voltage sets to the default value of 1000V.
d.
After setting, press the SAVE key to save the changes made or press the
ESC/RECALL MEM key to exit without saving and return to the previous screen
EN - 16
HT7051 - HT7051N

Publicidad

loading

Este manual también es adecuado para:

Ht7051n