ELECTROMAGNETIC COMPATIBILITY (EMC)
• This device should not be used adjacent to or stacked with other equipment and that if adjacent or stacked
use is necessary, this device should be observed to verify normal operation in the configuration in which
it will be used.
• Use of accessories other than those specified or provided by the manufacturer of this device could result
DE
in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result
in improper operation.
• Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the device, including cables
EN
specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
• When the operating environment is relatively dry, strong electromagnetic interference usually occurs. At
this time, the device may be affected as follows:
- the device stops output;
- the device turns off;
FR
- the device restarts;
The above phenomenon does not affect the basic safety and essential performance of the device, and
the user can use it according to the instruction. If you want to avoid the above phenomenon, please use it
according to the environment specified in the manual.
IT
Declaration - electromagnetic emission
ES
The device is intended for use in the electromagnetic environment specified below. The customer
or the user of device should assure that it is used in such an environment.
Emissions test
NL
RF emissions
CISPR 11
RU
PL
RF emissions
CISPR 11
SE
Harmonic emissions
IEC 61000-3-2
FI
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Table 1
Electromagnetic environment -
Compliance
guidance
The device uses RF energy
only for its internal function.
Therefore, its RF emissions are
Group 1
very low and are not likely to
cause any interference in nearby
electronic equipment.
The device is suitable for
domestic establishment and in
establishment directly connected
Class B
to the public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Not applicable
Not applicable
Table 2
Declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or
the user of device should assure that it is used in such an environment.
Immunity test
IEC 60601test level
Electrostatic
±8 kV contact
discharge (ESD)
±2 kV, ±4 kV, ±8 kV,
IEC 61000-4-2
±15 kV air
± 2 kV for power
Electrical fast
supply lines
transient/burst
± 1 kV for input/
IEC 61000-4-4
output lines
± 0.5kV, ± 1 kV
Surge
line(s) to lines
IEC 61000-4-5
± 0.5kV, ± 1 kV, ± 2
kV line(s) to earth
0 % UT; 0.5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
Voltage dips, short
interruptions and voltage
0 % UT; 1 cycle
variations on power
and70 % UT;
supply input lines
25/30 cycles
IEC 61000-4-11
Single phase: at 0°
0 % UT;
250/300 cycles
Power frequency
(50/60 Hz) magnetic
30 A/m
field IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Electromagnetic environment
Compliance level
- guidance
Floors should be wood,
concrete or ceramic tie.
±8 kV contact
If floors are covered with
±2 kV, ±4 kV, ±8
synthetic material, the relative
kV, ±15 kV air
humidity should be at least
30 %.
Mains power quality
should be that of a typical
Not applicable
commercial or hospital
environment.
Mains power quality
should be that of a typical
Not applicable
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the device requires
Not applicable
continued operation during
power mains interruptions,
it is recommended that the
device be powered from an
uninterruptible power supply
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
30 A/m
location in a typical
commercial or hospital
environment.