HHPC 6+, HHPC 3+, HHPC 2+ USER MANUAL
Operation
Option
Cycles
Mode
4
Configure the options for Location, Data display, Communications, Data export and Alarm
Setup as necessary.
Start a sample process with normal view
Remove the protective cap from the inlet and attach the correct probe (if necessary) before starting
a sample process.
1
Navigate to the Sample screen icon.
2
Push the
The instrument begins the sample process with the default or stored Sample Setup values.
While the instrument takes samples, the display shows the sample status, current sample
number and the sample time that remains
NOTE
If the Sample Mode is set to Volume, the value for the sample time that remains is an estimated
value based on the Volume in Sample Setup.
3
Let the instrument complete the sample process. To cancel the sample process, push the
button again.
SELECT
The instrument logs data to the data buffer. A status field shows errors (if any) that occurred in
the sample process.
NOTE
If the sample is manually terminated prior to completion, the data will not be saved.
16 / 26
Description
The total number of sample and hold intervals. Range: (0) to 999. When in Automatic
mode, the instrument stops after the last cycle is complete. A value of 0 makes the
unit run continuously until the user pushes the SELECT button.
Automatic: the instrument gets samples according to the stored parameters.
Manual: the instrument gets one sample and stops.
Beep: the instrument uses Automatic mode settings but ignores the count alarm
settings. The unit gives an audible beep for each particle counted.
button to start the sample process.
SELECT
(Sample
screen).
DOC026.97.80271, Edition 9