Problem
The display "Er:[No.]:11"
appears.
The display "Er:[No.]:12"
appears.
The display "Er:[No.]:16"
appears.
The Enter key
(send/print) is pressed
and the display
"Er:[No.]:71" appears.
The Enter key
(send/print) is pressed
and the display
"Er:[No.]:72" appears.
10.SERVICING/VERIFICATION
10.1 Information on servicing and verification
110 •
Cause / Solution
The scale has too high a load or too high a load in
one corner.
- Take the load off the scale or distribute the
weight more evenly
- Restart the scale
The scale has been switched on with too high a
load.
- Remove the load from the scale
- Restart the scale
The scale was caused to oscillate, the zero point
could not be determined.
- Restart the scale
No data transmission possible, wireless module
deactivated.
- Activate wireless module (see "Activating the
wireless module (SYS)" on page 102)
No data transmission possible, no wireless group
set up.
- Set up wireless group (see "Setting up a wire-
less group (Lrn)" on page 102)
We recommend having your device serviced prior to
verification.
NOTICE!
Incorrect measurements as a result of poor
servicing
►
Have servicing and repairs carried out
exclusively by seca Service or by an
authorized service partner.
►
You can find service partners in your area at
www.seca.com or by sending an e-mail to
service@seca.com.
Have an authorized technician perform verification
according to national legal regulations.