2.6.2 Manual emergency lowering
L
WARNING: Risk of injury – Adjust the manual emergency lowering for the weight
of the patient.
Please note that the manual emergency lowering is only possible when the patient is sitting
in the patient lift.
Pull force: 10 mm (0.39 in) for force from 80 à 100N.
3

Installation and adjustment

The instructions in this chapter are for the specialist dealer.
To find a service facility or specialist dealer near you, contact the nearest Vermeiren facility.
A list of Vermeiren facilities can be found on the last page.
L
WARNING: Risk of unsafe limitations - Use only the limitations described in this
manual.
3.1

Swapping the battery

We shall decline all liability for damage caused by the use of improper battery.
Do not use the battery at temperatures below +5°C (41°F) or above +50°C (122°F)
(the ideal is +20°C (68°F)).
If the battery is opened, all liability of the manufacturer shall be voided as well as any
claim.
The battery can be swapped by an integrated clip system.
Emergency lowering if the power fails or the batteries are
empty can be done using the red pull button (1) at the
lower end of the motor. This emergency lowering is
factory-set for a patient weight of 50 kg.
Standard calibration: 2500N, 15 mm/s to lower the
patiënt.
To swap the battery pull on the integrated clip
system. The plate will be lifted and the battery
can take off the holder (control box).
Page 13 
EN
Albatros
2010-12