Step 4:
Squeeze a drop of control solution onto a clean, dry, non-
absorbent surface. Do not apply control solution to the test
strip directly from the bottle. Replace the bottle cap on the
control solution bottle immediately after use.
DO NOT touch the Sampling End of the test strip with the bottle of
control solution. See illustration of Sampling End below.
Step 5:
Gently touch the drop of the control solution to the tip of the
mem
test strip. The meter will beep indicating it has received the
control solution.
Step 6:
mg /dL
ctl
AM
DAY
The screen will start to count down. After 6 seconds, the
PM
AVG
control test result will appear on the meter's display screen.
code
Step 7:
Compare the reading on the screen to the Low range
printed on the test strip bottle or its package. If the reading
falls outside the Low range printed on the test strip bottle
or its package, see Control Solution Trouble Shooting, on
page 29.
mem
NOTE:
• Control solution testing results will be stored into the meter memory and
indicated by ctl icon.
• Control solution testing results will not be used for calculating averages.
mg/dL
DAY
AVG
mg /dL
PM
mem
mg/dL
DAY
AVG
Control Solution Testing
PM
code
mol/dL
PM
mem
mg /dL
DAY
AVG
EvenCareG
U
s
e
t i
w
h
n i
3
m
o
n
t
h
s
a
t f
e
r
f
t s
r i
Low
Control Solu on
Suitable for Self-tes ng
o
r
e
b
f
o
e
r
h
t
e
e
p
x
a
r i
o i
t
n
d
a
e
t
I
V
D
1
5
o t
S
r
a
e
g
T
e
m
e
p
a r
u t
e r
PM
code
Low
53¡ª 79
High
135¡ª 203
Low
High
mmol/dL
mol/dL
PM
mem
mg /dL
DAY
AVG
27