Pump Test; Observe The Following During Measuring Mode With Pump - Dräger X-zone 5500 Instrucciones De Uso

Ocultar thumbs Ver también para X-zone 5500:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Operation with pump (optional)
8.2

Pump test

NOTICE
i
i
Dräger recommends inspection of the seals for
damage before attaching the holder (pump). If
damaged, the holder (pump) must be replaced.
The pump test is performed in the same way as for the
Dräger pump
X-am 1/2/5000. During the pump test, the default
setting acoustic alarm is automatically reduced to
80 dB (A).
The pump LED flashes red/
green, accompanied by a
signal tone.
Tightly close off intake
port or kink the sampling
hose at the beginning of
the hose for at least
2 seconds.
The pump LED lights red.
Unblock the intake flange/
sampling hose.
62
If the test was successful:
The pump LED lights green,
accompanied by an acoustic
acknowledgment signal.
If the test was not
successful:
The pump LED lights red,
accompanied by a continuous
tone.
The
pump
switches
automatically.
00733112.eps
Ending operation
1. Switch off the X-am 5x00 following the instructions for use.
The visual and the audible alarm will be activated for a
short time before the device switches off.
The X-zone 5500 is in STANDBY mode.
2. Unscrew the sampling hose or the Dräger probe from the
filter.
3. Remove the holder (pump) by releasing the lock.
The pump switches off.
4. Place the holder onto the housing.
02633112.eps
5. Fasten the lock at the holder.
8.3
Observe the following during measuring
mode with pump
Wait until the flushing time has elapsed.
Before every measurement, flush the Dräger sampling
hose or the Dräger probes with the air sample to be
measured.
A flushing phase is necessary to eliminate or minimise all
effects associated with the use of a sampling hose or a
probe, e. g. absorption in the hose, dead volume.
00833112.eps
The duration of the flushing phase depends on factors such
as type and concentration of the gas or vapour to be
measured, material, length, diameter, and age of the
sampling hose or probe. Generally, when using a sampling
hose (new, dry, clean), a typical flushing time of approx.
3 seconds is required for each metre. This flushing time
applies in addition to the sensor response time (see the
Instructions for Use for the gas detection instrument used).
off
Dräger X-zone 5500
00933112.eps
00933112.eps

Publicidad

Tabla de contenido
loading

Tabla de contenido