2. With central gas wall CO2 supply
As soon as the central gas wall CO2 pressure reaches 2 bars, the insufflators inform the user via :
· A message on the display in the warning area "input pressure".
· A yellow light displayed.
From, 2 bars insufflator is not allowed to re-start.
From, 10 bars insufflators will automatically switch in "bottle" mode for safety reasons.
Use only with central GAS WALL CO2 supply complying with European standards.
K. END OF OPERATION
Stop insufflation by pressing [F]; the STOP icon appears.
Disconnect the tubing from the trocar and the insufflator immediately to prevent any liquid or gas from flowing back to the device.
The used tubing must be disposed of in an appropriate container after use.
The total volume of gas insufflated during the operation appears; reset the gas meter to zero if another operation is planned (see chapter VII – C).
Close the CO
gas cylinder.
2