Accuracy is ± [% reading (R) + number of digits (D)].
{Accuracy : ''n %R + n D'' means ''n % of the reading + n digits'' }
5.2.1. DC voltages
Selector switch
position
ADP
DC
V
DC
(∗) Acceptance permanent max. voltage
Number of points :
Range selection :
Common mode rejection (500 mV range) :
Serial mode rejection :
Intermittent signal sounds and « OL » on the display in case of range overpassing.
5.2.2. AC voltages (AC and AC+DC)
Selector switch
position
V
LOW Z
ADP
AC+DC
V
AC+DC
(∗) Acceptance permanent max. voltage
Specifications applicable from 5 % to 100 % of the range for a sinusoidal signal in band
40 Hz to 1 kHz.
Number of points :
Range selection :
Common mode rejection :
Additional error according to crest factor :
Intermittent signal sounds and « OL » on the display in case of range overpassing.
5000-count portable digital multimeter, with current input
Ranges
Accuracy
500 mV
DC
5 V
DC
50 V
0,3 %R + 2 D
DC
500 V
DC
1000 V
DC
Ranges
Accuracy
40 Hz to 1 kHz
5 V
AC
50 V
1 %R + 2 D
AC
500 V
AC
600 V
AC
500 mV
1.5 %R + 3 D
AC+DC
5 V
AC+DC
50 V
1.5 %R + 2 D
AC+DC
500 V
AC+DC
750 V
AC+DC
Input impedance
Protection
± 600 V
10 MΩ
11 MΩ
± 1100 V
10 MΩ
5 000
automatic or manual for the 5 V, 50 V,
500 V, 1000 V ranges
at 50 and 60 Hz, better than 120 dB
at 50 and 60 Hz, better than 60 dB
Input
Protection
impedance
± 600 V
500 kΩ
± 600 V
10 MΩ // 100 pF
11 MΩ // 100 pF
± 1100 V
10 MΩ // 100 pF
5 000
automatic or manual for the 5 V, 50 V,
500 V, 600 or 750 V ranges
at 50 and 60 Hz, better than 60 dB
1 % for a crest factor of 1.5 to 2
4 % for a crest factor of 2 to 3
Resolution
(∗)
0.1 mV
RMS
DC
1 mV
DC
10 mV
DC
PEAK
100 mV
DC
1 V
DC
Resolution
(∗)
1 mV
AC
10 mV
RMS
AC
100 mV
AC
1 V
AC
0.1 mV
AC+DC
RMS
1 mV
AC+DC
10 mV
PEAK
AC+DC
100 mV
AC+DC
1 V
AC+DC
23