Instrument settings

The following parameters can be changed on an instrument:
Name
Password
1)
Operating signal LED
1)
Horn operating signal
Switch-off mode
2)
Shift length (TWA)
Short-term exposure level (STEL)
1)
At least one of the two operating signals must be switched on.
2)
Corresponds to averaging time and is used to calculate the TWA exposure value.
3)
Interpretation only if the sensor is designed for this.
4)
Corresponds to averaging time and is used to calculate the STEL exposure value.
Sensor settings
The following parameters can be changed on the sensors:
Name
Alarm threshold A1 (in measuring unit)
Alarm threshold A2 (in measuring unit)
1)
Interpretation type
Alarm threshold STEL
1)
(in measuring unit)
Alarm threshold TWA
1)
(in measuring unit)
1)
Interpretation only if the sensor is designed for this.
Check of parameters
To ensure that the values were correctly transferred to the gas detection
instrument:
Press the data from X-am 1/2/5x00 in Dräger CC Vision.
Check parameters.
Range
Numerical range (3-digit)
Yes / No
Yes / No
"switch-off allowed" or
"switch-off prohibited" or
"switch-off prohibited at A2"
60 - 14400 (in minutes)
(Setting for exposure alarm)
3) 4)
0 - 15 (in minutes)
(Setting for exposure alarm)
Range
0 - A2
A1 – measuring range end value
Inactive, TWA, STEL, TWA+STEL
0 – measuring range end value
0 – measuring range end value
Operation
Preparations for operation
Before using the instrument for the first time, insert a charged NiMH T4
power pack or batteries approved by Dräger see "Replacing the batteries /
rechargeable batteries" on Page 20.
The instrument is now ready for operation.
To reduce the risk of ignition of a flammable or explosive atmosphere, strictly
adhere to the following warning statements:
Only use power pack types ABT 01xx, HBT 00xx or HBT 01xx. See the marking
on the rechargeable battery for permitted rechargeable batteries and the
corresponding temperature class.
Substitution of components may impair intrinsic safety.
Switching on the instrument
Hold down the [ OK ] key for approx. 3 seconds until the countdown » 3 . 2 . 1 «
shown on the display has elapsed.
All the display segments, including the visual, audible and vibration alarms,
are activated for a short time.
The software version is displayed.
The instrument performs a self-test.
The sensor that is up next for adjustment is displayed with the remaining
days until the next adjustment, e.g. » ch
The time until the bump test interval elapses is displayed in days, e.g.» bt 2 «.
All A1 and A2 alarm thresholds and if applicable »
1)
(STEL)
for all toxic gases (e.g. H
During the sensor warm-up phase:
The display for the measured value flashes
The special symbol »
« is displayed.
No alarms are issued during the warm-up phase.
The red LEDs flash.
The gas detector is ready to measure when the measured values no
longer flash and the red LEDs are no longer illuminated. The special
symbol »
« may continue to be displayed if corresponding warnings
(e.g. not yet ready for calibration) are active (to view the warnings, see
the technical manual).
1)
Only when activated in the instrument configuration. Delivery condition: not activated.
WARNING
%LEL CAL 123 «.
4
1)
« (TWA)
and »
S or CO) are displayed consecutively.
2
«
17