46
5.3
Command list for RS communication
The commands have the following form:
Adresse
2-digits aa,
Command
Parameter
Each command must be entered separately and ended by <CR> <LF>. All replies will be returned to the
computer only upon completion of the respective action.
Command
Description
aaBF
Fill burette
aaBH14
Number of steps up 1 to 10000
aaBI3456
Step increment 1 to 10000
aaBN
Reset of ml display to zero
aaBR145
Number of steps down 1 to 10000
aaBP
Inquire piston position in scale units
aaBV
Inquire burette volume
aaDA13.56
Carry out dosing with addition of the volume
aaDO123.4
Dose the specified quantity in ml
aaDB123.4
Dose quantity in ml without filling and zero
aaDM34.56
Carry out dosing (ml) with sensor supervision
aaGDH11,3
Speed for dosing in ml/h
aaGDM20,3
Speed for dosing in ml/min
aaGF20
Filling time in seconds (min 20s)
aaOA1
Switch off specified output 1
aaOE1
Switch on output 1
aaOJ
Switch on output
aaON
Switch off output
aaRH
Identification request
aaRA
Inquire unit size
aaRS
Report status
aaRC
Inquire "last command"
aaVD123.4
Dose the specified quantity in ml
aaVE
Inquire software version no.
aaVF
Valve in 'fill' position
aaVN
Valve in 'dose' position
aaVO
Volume inquiry from display
aaSEEPROM
Reset EEPROM to ex-factory values
aaSR
Stop current function
Example: Dose 12.58 ml using the TITRONIC
"03DO12.58 <CR> <LF>". Reply from the piston burette upon completion of the dosing operation: "03Y
<CR> <LF>".
Connections and data transfer
e.g.: 01
e.g.: BP
e.g.: 14 <CR> <LF>
®
110 plus Piston Burette having the address "03":
Reply
aaY
aaY
aaY
aaY
aaY
aaBP=1234
aaV=12.34
aaY
aaY
aaY
aaY
aaY
aaY
aaY
aaY
aaY
aaY
aaY
aaIdent T200
aaAUFSATZ:20
aaStatus:"text"
aa"last comm."
aaY
aaVersion:9049
aaY
aaY
aaV=12.345ml
aaY
aaY