• After each Operation, be sure to lubricate and sterilize the handpiece as soon as it's cleaned. Blood coagulation can
cause corrosion and rusting. However, do not lubricate the Micromotor. Oil could generate excessive heat and cause damage.
• The Control Unit can be cleaned with a moist cloth. Disconnect the power supply before cleaning. The Control Unit
and the Foot Control cannot be sterilized by any method.
• When the system is very frequently used, consider the maintenance of a small stock of replaceable parts.
• Do not disconnect the Motor Cord from the motor.
• Do not wipe with or, clean or immerse in, high acid water or sterilizing solutions.
• Follow the table below for the motion time of motor and handpiece. Continuous
use over a long time with high torque may cause the overheating of the
handpiece, which in turn may result in an accident.
• Applied parts for patient and/or operator is Handpiece.
• Do not fixate the eyes of LED light.
• If any abnormalities (dark, do not light on or flashing) are found for LED, stop
using the it immediately and contact dealer.
• Using the Micromotor with light 'ON' for more than 3 minutes continuously,
motor will become hot. (The maximum surface temperature might reach 45˚
C by the situation used) In such cases, stop using the Micromotor till the light
part cools down, or use the Micromotor without turning on the light.
• Grounding reliability can only be achieved when the equipment is connected to
an equipment receptacle marked "Hospital Only" or "Hospital Grade".
• For safety's sake, install the Control Unit in a place where the AC POWER
Cord can be easily removed. (It is possible to disconnect the control unit from the power source by removing the AC
POWER Cord.)
• Turn off the Main Power Switch after each use.
• For service requirements and spare parts contact dealer.
• Use of NSK genuine pre-sterilized, disposable Irrigation Tube is recommended.
• Total recording time is about 100 minutes. (Surgic Pro + )
• Do not insert a USB flash drive during operation.
• Do not remove USB flash drive while LED lamp is on or blinking.
• Do not use data stored in USB flash drive for diagnostic purposes.
• Do not use a USB flash drive other than USB 2.0.
• This device is not guaranteed to work with all USB flash drive. It may not work with USB flash drive with security
functions or those requiring a great deal of electricity.
No moisture condensation in the Control Unit.
Use at outside of these limits may cause malfunction.