Bench Alignment - Ottobock 1S101 SACH+ Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 37
12) Wipe off adhesive that comes out of the joint using a cloth.
13) Remove any remnants of the adhesive using a suitable solvent.
14) Allow the adhesive to cure (curing time: 16 h).
5.3 Alignment
NOTICE
Grinding the prosthetic foot
Premature wear resulting from damage to the prosthetic foot
► Do not grind the prosthetic foot.

5.3.1 Bench Alignment

TT bench alignment
Required materials: 662M4 goniometer, 743S12 heel height measuring apparatus,
743A80 50:50 gauge, alignment apparatus (e.g. 743L200 L.A.S.A.R. Assembly or
743A200 PROS.A. Assembly)
Perform the assembly and alignment of the prosthetic components in the alignment
apparatus according to the following specifications:
Sagittal plane
Heel height: Effective heel height (shoe heel height - sole thickness of fore­
   
foot) + 5 mm
Exterior foot rotation: approx. 5°
   
   
Anterior placement, middle of the prosthetic foot to the alignment reference
line: 30 mm
   
Connect the prosthetic foot and prosthetic socket using the chosen adapters.
The instructions for use of the adapters must be observed.
   
Determine the centre of the prosthetic socket with the 50:50 gauge. Align the
prosthetic socket centrally to the alignment reference line.
Socket flexion: Individual residual limb flexion + 5°
Frontal plane
Alignment reference line of prosthetic foot: between big toe and second
   
toe
Alignment reference line of prosthetic socket: along the lateral patella
edge
   
Observe the abduction or adduction position.
16
Bench alignment process

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

1s102 sach+1s103 sach+1s1011s1021s103 sach serie

Tabla de contenido