Use
If the gas monitor is turned off, the values for the peak concentration, TWA
and STEL are deleted.
4.3

During operation

!
WARNING
Danger to life and/or risk of explosion!
The following alarms may indicate a danger to life:
A2 alarm
STEL or TWA alarm
Device error
► Immediately leave the hazard area.
The continuous operation of the gas monitor is indicated by an optical and/or
acoustic operation signal that is emitted in a 60-second cycle (configurable using
the Dräger CC-Vision PC software).
The acoustic operation signal must be switched on for measurements in
accordance with EN 45544 (CO, H
If the allowable measuring range is exceeded or a negative zero drift occurs, the
following message appears in the display:
(negative drift).
The measuring channels do not need to be checked after a short-term over
range (up to one hour) by the EC measuring channels (this does not apply if
using DrägerSensor XXS CO H
In the event of an alarm, the corresponding displays, the visual, audible and
additionally the vibration alarm, are activated, see section 6.2 on page 25.
Push the [▼] button to light up the display.
If the gas detector is used for offshore applications, a distance of 5 m to a
compass must be complied with.
22
S) or in accordance with EN 50104 (O
2
(concentration too high) or
-CP).
2
).
2