Maintenance And Repair; Specifications - Pari Trek S Instrucciones De Uso

Ocultar thumbs Ver también para Trek S:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

MAINTENANCE AND REPAIR: Never operate the PARI TREK® S Compressor if it is
improperly or incompletely assembled, or if it is damaged. Use only original PARI parts,
accessories, tubing and attachments authorized by PARI Service Center for the PARI
TREK® S Compressor Nebulizer System. Use of other parts, accessories, tubing and
attachments may damage the PARI TREK® S Compressor unit or cause improper treatment.
Do not open or disassemble the PARI TREK® S Compressor unit at any time. This action
will void the warranty. Repairs, service and maintenance should be carried out only by
an authorized PARI Service Center.
NOTE: There are no user serviceable parts inside the PARI TREK® S Compressor or Battery
housing. Service by anyone other than an authorized PARI Service Center voids the warranty.
CLEANING THE RECHARGEABLE BATTERY: Use a clean, dry cloth to wipe the exterior of
battery. Be careful not to get liquids onto the Charge Receptacle or the Battery
Contacts since damage could occur.
PARI TREK® S COMPRESSOR NEBULIZER SYSTEM SPECIFICATIONS
Weight: (Compressor only).......................................................0.8 lb (360 grams)
With Rechargeable Battery:.......................................................1.2 lbs (520 grams)
Dimensions: (Compressor only)..................................................4.5" x 3.5" x 1.75"
Certification:.............................................IEC/EN 60601-1 and IEC/EN 60601-1-2
Compressor Performance:
Maximum Pressure of Compressor:................................................40 PSI (2.75 Bar)
Maximum Liter Flow of Compressor:.............................................................8 LPM
Operating Pressure of Compressor¹:.............................................14.5 PSI (1.0 Bar)
Operating Liter Flow of Compressor¹:.........................................................4.2 LPM
Ambient Operation:
Temperature Range:...................................................50ºF to 104ºF (10ºC to 40ºC)
Transport/Storage:
Ambient Temperature:.............................................-13ºF to 158ºF (-25ºC to 70ºC)
Humidity:....................................................................................10% to 95%
Atmospheric Pressure:............................................................500 hPa - 1060 hPa
Compressor Electrical Requirements:.........................Input 12 VDC 1.67 A 18 watts
¹ Measured with PARI LC® Sprint or LC® Plus Reusable Nebulizer attached.