6. Batteries substitution.
SMART TENS is supplied by 4 AAA Size 1.5V Alkaline batteries. Display will show low
battery indicator
appearing only when the battery is too low, and there is no indication on the
display regarding the batteries status during the treatment.
To substitute the batteries follow these steps:
1. Unlock the belt clip on the back side of SMART TENS.
2. Open battery compartment.
3. Extract the batteries from the compartment.
4. Insert 4 new AAA Size batteries (we recommend to always substitute the 4 batteries
at the same time) respecting the polarity as reported on the battery compartment.
5. Close battery compartment.
6. Lock the belt clip.
7. Check the functioning of SMART TENS pressing ON/OFF
chapter "Preliminary instruction")
7. SMART TENS maintenance and storage.
We recommend that SMART TENS is switched off and the cables are removed from
the connectors at the end of every session. SMART TENS should be kept in the carry
bag supplied, together with the rest of the equipment supplied with it, and carefully
stored on a secure surface and protected from situations listed in the Warnings
paragraph.
SMART TENS should be stored in the following conditions:
Temperature:
Maximum relative humidity:
Atmospheric pressure:
Cleaning: only clean the device using a dry cloth.
We recommend a functional test carried out by the manufacturer every 24 months.
The manufacturer retains that the SMART TENS device cannot be repaired by
personnel external to its own company. Any work of this nature carried out by personnel
not authorised by the manufacturer will be classed as tampering with the device, freeing
the manufacturer from responsibility for the warranty and hazards that the operator or
user may be exposed to.
N.B. disconnect the cables before placing the electronic stimulator in the bag. If not, the
cables will bend excessively close to the connectors. This may damage the cables.
I.A.C.E.R. S.r.l.
only when the batteries are low: this is a graphic icon
from -5°C to +40°C
from 10% to 93%
from 700 hPa to 1060 hPa
32
button (see the
MNPG45-02