Technical Details - Enraf Nonius Crosswalker Instrucciones De Manejo

Ocultar thumbs Ver también para Crosswalker:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

16 Technical details

Mains voltage
Power supply
Frequency
Permitted voltage variation
Power consumption
Braking principle
Load
Accuracy of rpm
Heart rate measurement accuracy : heart rate measurement with Polar®, +/-2% or 1 beat per minutes
Min. adjustable power
Medical class
Accuracy class
Electrical safety
Dimensions
Weight
Max. patient weight
Display
Languages
Heart rate monitoring
Exercise modes
Parameters displayed
during training
Operating conditions:
Temperature
Humidity
Air pressure
Storage and transport conditions:
Temperature
Humidity
Air pressure
Program capacity
Fixed programs
Installation of the Crosswalker with chip-card function on the EN-Track software.
You must have at least version 6.5 of the EN-Track software to be able to operate the machine in a chip-
controlled manner.
The patient shall receive instructions to stop training when prescribed environmental conditions are not
met.
Technical changes reserved.
44
: 230 V
: via power cord (230 V EUR)
: 50 Hz
: +/- 10%
: approx. 40VA
: process-controlled eddy-current brake
: 20-400 W (resolution: 5 W increments) rpm-independently from
35 – 100 rpm
: 6 – 60 Nm (resolution: 0.5 Nm increments) rpm-independently
from 35 – 100 rpm
: rpm, revs +/- 2 min¯¹
: power in Watts, +/- 10% or 5 W. (depending on which value
is the larger) See also the range in Fig. 20.
(depending on which value is the larger)
This definition assumes constant conditions.
Heart rate measurement with ear clip +/- 5%
This definition assumes constant conditions.
: 20 W
: class IIa
: class SA (EN 957)
: Class II type B (3446668) or BF (3446673)
(according to IEC 60601-1)
: 190 cm x 177 cm x 55 cm (l x h x w)
: approx. 80 kg
: 150 kg / 330 lbs
: LCD with backlighting
: English, Dutch, German
: Polar® chest strap and receiver and hand/pulse measurement
(options)
Ear sensor prepared.
: constant torque, constant power, constant heart rate
: time, heart rate, speed, distance, work, power and resistance
: 18 - 40 °C
: 10 - 85% at 25 - 40 °C, non-condensing
: 560 - 1500 hPa
: 14 - 40 °C
: 30 - 95% at 25 - 40 °C, non-condensing
: 560 - 1500 hPa
: 12 fixed exercise profiles

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Cruiser

Tabla de contenido