Specifications
HOLD input
Galv. separated (OPTO coupler)
Function
Switches Transmitter to HOLD mode
Switching voltage
0 ... 2 V (AC/DC)
10 ... 30 V (AC/DC)
CONTROL input
Galv. separated (OPTO coupler)
Function
Switch-over to second parameter set
Switching voltage
0 ... 2 V (AC/DC)
10 ... 30 V (AC/DC)
Output 1
0/4 to 20 mA, max. 10 V, floating
(galv. connected to output 2)
Measured variable
Conductivity, concentration or salinity
*)
Characteristic
Linear or logarithmic
Overrange
*)
22 mA in the case of error messages
Output filter
*)
(attenuation) Low-pass, filter time constant 0 ... 120 s
Measurement error
1)
< 0.3 % current value + 0.05 mA
Start/end of scale
As desired within range
Min. span
LIN: 5 % of selected range
LOG: 1 decade
Output 2
0/4 ... 20 mA, max. 10 V, floating
(galv. connected to output 1)
Process variable
Temperature
Overrange
*)
22 mA in the case of temp error messages
Output filter
*)
Low-pass, filter time constant 0 ... 120 s
Meas. error
< 0.3 % current value + 0.05 mA
1)
Start/end of scale
*)
-20 to +200 °C / -4 ... +392 °F
Adm. span
20 ... 320 K (36 to 608 °F)
Alarm contact
Relay contact, floating
Contact ratings
AC< 250 V / < 3 A / < 750 VA
DC< 30 V / < 3 A / < 90 W
Contact response
N/C (fail-safe type)
Alarm delay
0000 ... 0600 s
100
Hold inactive
Hold active
Parameter set 1
Parameter set 2
Cond Ind 7100 e
Limit values
Output via relay contacts R1, R2
(see PID process controller)
Contacts R1, R2 floating but inter-connected
Contact ratings
*)
AC< 250 V / < 3 A / < 750 VA
DC< 30 V / < 3 A / < 90 W
Contact response
N/O or N/C
*)
Delay
*)
0000 ... 9999 s
Switching points
*)
As desired within range
Hysteresis
0 ... 50 % full scale
*)
PID process controller Output via relay contacts R1, R2
(see limit values)
Setpoint
*)
As desired within range
Neutral zone
*)
As desired within range
Proportional action
Controller gain K
*)
Integral action
*)
Reset time
(0000 s = no integral action)
Derivative action
*)
Rate time
(0000 s = no derivative action)
Controller type
Pulse length or pulse frequency controller
*)
Pulse period
*)
0001 ... 0600 s, min. ON time 0.5 s
(pulse length controller)
Max. pulse frequency
0001 ... 0180 min
*)
(pulse frequency controller)
Cleaning function / Parameter set 2
*)
Clean / PSEt2
Relay contact, floating,
for controlling a rinsing probe or
signaling that 2nd parameter set is active
Contact ratings
AC< 250 V / < 3 A / < 750 VA
DC< 30 V / < 3 A / < 90 W
Contact response
N/O when signaling parameter set 2
N/O or N/C when used as cleaning contact
Rinsing interval
000.0 ... 999.9 h
*)
(000.0 h = cleaning function switched off)
Rinse duration
*)
0000 ... 1999 s
: 0010 ... 9999 %
C
T
: 0000 ... 9999 s
R
T
: 0000 ... 9999 s
D
-1
*)
101