5.1.16 A ppliance dimensions:
(L x W x H) = 158 x 76 x 38 mm without rubber protection frame
(L x W x H) = 164 x 82 x 44 mm with rubber protection frame
Appliance weight:
240 g (MM 8), 265 g (MM 9/ 10) without rubber protection frame
340 g (MM 8), 365 g (MM 9/ 10) with rubber protection frame
5.1.17 T he safety measuring cables are designed in 4 mm plug-in type technol-
ogy. The safety measuring cables supplied are expressly suited for the
rated voltage
5.1.18 T he BENNING MM 8/ 9/ 10 is protected by a rubber protection frame
against mechanical damage. The rubber protection frame allows the
B ENNING MM 8/ 9/ 10 to be set up or hung up during the measure-
ment.
5.1.19 T he BENNING MM 10 features at its head an optical infrared RS-232
interface for accepting the IR adapter on the RS-232 cable. The
optical infrared RS-232 interface is used for the galvanic separation of
the measuring signal and the transmission of measuring data to a PC/
laptop.
6. Environmentconditions
-
T he BENNING MM 8/ 9/ 10 is intended for making measurements in dry
environment.
-
M aximum barometric elevation for making measurements: 2000 m,
-
Overvoltage category/ Siting category: IEC 60664-1/ IEC 61010-1 → 600 V
category IV; 1000 V category III,
-
C ontamination class: 2,
-
Protection Class: IP 30 (DIN VDE 0470-1 IEC/ EN 60529)
I P 30 means: Protection against access to dangerous parts and protection
against solid impurities of a diameter > 2.5 mm, (3 - first index). No protec-
tion against water, (0 - second index).
-
O perating temperature and relative humidity:
F or operating temperature from 0 °C to 30 °C: relative humidity less than 80 %
F or operating temperature from 31 °C to 40 °C: relative humidity less than
75 %
F or operating temperature from 41 °C to 50 °C: relative humidity less than 45 %
-
S torage temperature: The BENNING MM 8/ 9/ 10 can be stored at any
temperature in the range from - 20 °C to + 60 °C (relative humidity from 0
to 80 %). The battery should be taken out of the instrument for storage.
7. Electricalspecifications
Note: The measuring precision is specified as the sum of
-
a relative fraction of the measured value and
- a number of digits (counting steps of the least significant digit).
This specified measuring precision is valid for temperatures in the range from
18 °C to 28 °C and relative humidity less than 80 %.
7. Directvoltageranges
The input resistance amounts to 10 MΩ (within the 400 mV range 1 GΩ).
Measuring
Resolution
range
600 mV
6 V
60 V
600 V
100 mV
1000 V
7.2 Alternatingvoltageranges
The input resistance amounts to 10 MΩ parallel 100 pF.
Measuring
Resolution
range
600 mV
6 V
60 V
600 V
100 mV
750 V
11/ 2006
and the rated current of the BENNING MM 8/ 9/ 10.
100 µV
± (0.5 % of the measured value + 2 digits)
1 mV
± (0.5 % of the measured value + 2 digits)
10 mV
± (0.5 % of the measured value + 2 digits)
± (0.5 % of the measured value + 2 digits)
1 V
± (0.5 % of the measured value + 2 digits)
Meas. precision
100 µV
± (0.9 % of the measured value + 5 digits)
1 mV
± (0.9 % of the measured value + 5 digits)
10 mV
± (0.9 % of the measured value + 5 digits)
± (0.9 % of the measured value + 5 digits)
1 V
± (0.9 % of the measured value + 5 digits)
B ENNINGMM8/9/10
Meas. precision
within the frequency
*1
range 50 Hz - 500 Hz
Overload
protection
1000 V
DC
1000 V
DC
1000 V
DC
1000 V
DC
1000 V
DC
Overload
protection
750 V
*2
eff
750 V
*2
eff
750 V
*2
eff
750 V
*2
eff
750 V
*2
eff