Sample measurement
Set reference temperature:
When
appears and the current reference temperature blinks, use the
or
key to toggle
between 25 °C and 20 °C. Press the
key to confirm your selection.
Set temperature correction method:
When
appears and the current setting blinks, use the
or
key to toggle between
linear and nLF (nonlinear factor compensation). Press the
key to confirm your selection.
If you choose the linear temperature correction method, the current temperature coefficient appears. Use
the
and
keys to increase or decrease the value. Press
to confirm your setting.
Set MTC temperature:
When
appears, use the
and
keys to increase or decrease the value of the temperature
for your sample. Press the
key to confirm your setting. The default setting is 25 °C.
Set TDS factor:
When the current TDS factor appears, use the
and
keys to increase or decrease the value.
Press the
key to confirm your setting.
5.3.3

Data output

If a PC or printer is connected, every endpointed reading is sent to the PC or printed via the RS232
interface.
By pressing and holding the
key,
appears. The meter sends out a reading every second
until it endpoints.
METTLER TOLEDO SevenEasy™ conductivity
9