5.6.REINITIALISING THE INSTRUMENT
To reinitialise all the test parameters, please follow the procedure below:
1. Switch off the instrument by setting the ON/OFF key to the OFF position.
2. Press the SET UP key and keep it pressed while switching on the instrument.
Machine Tester followed by Hard Reset are displayed for a few moments, then the main menu opens.
NOTES : The reinitialised values of the adjustable parameters correspond to the requirements of standard EN 60204.
List of the test parameters per function, with their initial value.
Function
U
N
HV
Imax
t
∆U
∆U
t
R
AC
R
10A
max
t
U
N
R
min
MΩ
t
∆t
Syst
t
Toutes
Baud
fonctions
To reinitialise all the test parameters and erase all stored data, please follow the procedure below :
1. Switch off the instrument by setting the ON/OFF key to the OFF position.
2. Press the SET UP and MEM/MR keys and keep them pressed while switching on the instrument.
Machine Tester then Hard Reset followed by Clear Memory are displayed for a few moments, then the main menu opens.
NOTES :
The reinitialised values of the adjustable parameters correspond to the previous table, in accordance with
standard EN 60204.
5.7.
CONFIGURATION
5.7.1. Display contrast
If the display is not sufficiently readable (screen too dark or lettering intensity too low), the display contrast must be reinitialised.
Parameter
Test voltage
Tripping current
Timer
Max permitted
max
voltage drop
Timer
Max permitted
resistance
Timer
Test voltage
min. allowed
insulation
resistance
Timer
Measuring
system
Timer
ON/OFF (M/A)
Baud rate
Buzzer
Contrast
Range of adjustment
or possible values
1, 1.25, 1.5 kV AC
1, 3, 5, 10, 20, 30, 40, 50, 60, 70,
80, 90, 100, 150, 200, 250, 300,
333, 350, 400, 450, 500 mA
1s...1min...5 min (in steps of 1 s and 10 s)
5.0 V (0.50 mm
2
), 5.0 V (0.75 mm
3.3 V (1.0 mm
2
), 2.6 V (1.5 mm
1.9 V (2.5 mm
2
), 1.4 V (4.0 mm
1.0 V ≥ 6.0 mm
1 s...15 s
10 mΩ...1000 mΩ (in steps of 10 mΩ)
or
mΩ (no limit)
*
1s...15 s
500, 1000 V DC
0.2 MΩ...1 MΩ (in steps of 0.1 MΩ)
1 MΩ...10 MΩ (in steps of 1 MΩ)
10 MΩ...500 MΩ (in steps of 10 MΩ)
Ω (no limit)
or
*
2 s...1min...10 min (in steps of 1 s and 10 s)
2-point (1 s), 4-point (5 s)
ON ou OFF
300, 600, 1200, 2400, 4800
ON/OFF
0%...100% (in steps of 2%)
44
2
),
2
),
3.3 V (1.0 mm
2
),
2
Initial value
1 kV
500 mA
1 s
2
)
10 s
mΩ
*
10 s
500 V DC
1 MΩ
15 s
2-point (1 s)
ON
4800
ON
50%