The procedure to follow is as follows :
1. Connect the remote control pedal in accordance with fig 41.
2. Record a first measurement at the desired memory location (memory number and machine number) with the keys on
the keyboard (see § 5.2).
3. Carry out the next test by pressing the START/STOP pedal.
4. Save the results by pressing the SAVE pedal.
The following window (fig. 42) is displayed for a moment (if for example an HT test
has been carried out) :
Then the main menu is displayed again. Continue the measurements.
Technical specifications of the pedal :
Cable length : 10 m
Commands
: START/STOP, SAVE
Case
: metallic
Weight
: 2 kg
Dimensions (W x H x D) : 300 x 175 x 55 mm

5.10. USE OF WARNING LAMPS (STANDARD VDE 104)

The warning lamps are used to warn the user and other persons who may be present on the sites of the tests of the presence of
dangerous voltages during dielectric tests.
!
Meaning of the colours :
- The red light (TEST) lit means the presence of a dangerous voltage at the terminals of the dielectric test. Please take care
during handling of the test probes (guns) ! DANGER PRESENT !
- The green light (READY) lit means that the instrument is powered up and ready for a measurement; there is not a dangerous
voltage at the terminals of the dielectric test. POTENTIAL DANGER !
Technical specifications of the warning lamps:
Cable length : 2 m
Bulbs
: 12-15V / 4W
Case
: plastic
Weight
: 0.3 kg
Dimensions (W x H x D) : 200 x 110 x 95 mm
!
NOTE : If no lamp lights up when the rotary switch is in the HV position, stop the measurements immediately and check the
connection of the warning lamps as well as the bulbs.
230V / 50Hz
500VA
REMOTE
WARNING LAMP 12V~
RS232C
INSULATION (M ) W
D
DISCHARGE TIME ( t)
500 - 1000 V
+
TEST
600V
600V 250V CAT III
ON
MAINS
OFF
TEST
D
TEST ON
CONTINUITY 12V/>10A
( U,R
10A
)
MEM/MR
PRINT/RS
SET-UP
C1
P1
P2
C2
DIELECTRIC (HV)
CHAUVIN
C.A 6121
ARNOUX
START / STOP
1 - 1.25 - 1.5 kV
MACHINE TESTER
Fig.43. Connection of the warning lamps to the C.A 6121 MACHINE TESTER
READY
TEST
Instrument connector
(female) type MAB 5100
2 3 4
1
5
DIN 41524 (front view)
and internal instrument
circuitry
front view
47
HV
1000V
500mA
20.3
U:1.08 kV
mA
No:028 FR
MACH:127
MEM
No
MACH
Fig.42. IMemory and machine
number information