NOTE
·
It is suggested to keep the packaging in case you need to ship the scanner for technical
assistance.
6.2

Hardware setup

NOTE
·
The scanner must be placed on a solid, rigid and skid-proof surface (laboratory desks,
racks) so that a horizontal safe position is ensured.
·
The scanner should not be placed directly in the sun light or close to a spot light
source.
·
The scanner should not be placed in a position in which it can come into contact with
liquids or fluid leaks.
·
If the scanner is installed in a very dusty environment protect it as much as possible.
When not used always close the front door. In addition a dust cover is provided in
order to secure an optimal protection.
·
The room temperature should not be less than 10 °C (50 °F) or more than 40 °C
(104 °F).
·
The humidity should be less than 80 %.
CAUTION
·
The table should have a load bearing capacity of at least 40 kg / 88 lbs.
·
Take care not to pinch your fingers when placing the scanner on the table.
·
Air inlet and outlet openings in the housing of the unit (rear and side surfaces) must
not be covered or blocked. The rear and side of the unit must be at least 100 mm /
3.94 inch away from the wall to avoid blocking the air outlet of the built-in fan, it
means it is required a free area (w x d: 550 x 550 mm / 21.65 x 21.65 inch).
·
The scanner needs to be positioned with easy accessibility to access the main socket
and disconnection from mains is possible, if needed.
·
Do not initialize the scanner prior to acclimatization.
·
Do not put the device in operation while the dust cover is in place.
W19641_REF_66069511_13_QSG_02GB_caraScan4.0_V02