Virtual SVV™ Instructions for Use – EN
Step 6
Replace the securing screw gently to avoid any unintentional opening of the compartment.
NOTICE
After inserting new batteries it is important that the charging process is continued until the yellow LED
charge indicator extinguishes. Do not disconnect the base station from the AC outlet during this
procedure.
How to clean Interacoustics products
Always remove the face cushion before cleaning the virtual reality goggle. Wipe the surface of the virtual
reality goggle and handheld remote with a soft, clean cloth or tissue. The cloth may be slightly dampened
with water, mild cleaning liquid (e.g. dish washing cleaner or similar).
•
Before cleaning always switch off and disconnect from AC outlet
•
Do not use solvents or aggressive cleaning liquids or pastes
•
Do not use disinfectant sprays
•
Do not use saturated cloths or tissue
•
Do not soak or immerse any of the connectors in liquid. Avoid getting the device wet
•
Do not autoclave, sterilize or immerse the instrument or accessory in any fluid
•
Use a soft cloth lightly dampened with cleaning solution to clean all exposed surfaces
•
Do not use hard or pointed objects to clean any part of the instrument or accessory
•
Face cushions are single use components. They must be exchanged before each new patient
Recommended cleaning and disinfection solutions:
•
Warm water with mild, nonabrasive cleaning solution (soap)
•
70% isopropyl alcohol
NOTICE
Cleaning the head strap
The elastic head strap can be machine-washed at 40° C with regular washing powder. However, do not
dry in the dryer. Note that the elasticity can deteriorate after 4-5 wash cycles. The head strap must be
discarded and exchanged at this point in order to ensure a comfortable and tight fit during testing.
Page 34