If you wish to change programs or adjust your volume and/or sensitivity setting, you
will need to attach the Neptune Connect. The sound processor does not need to be
turned off prior to attaching the Neptune Connect. Once the desired changes have
been made, you may remove the Neptune Connect by pressing the release latch,
and the sound processor will maintain the new settings. Attach your Neptune Color
Cover when the Neptune Connect is not in use. If you choose to leave the Neptune
Connect attached during use, any changes to volume and sensitivity take effect
immediately.
Sync Light
The Sync Light, located between the volume and sensitivity dials, is a safety feature
designed to ensure the Neptune Connect is using the same settings as the sound
processor. If you remove and re-attach the Neptune Connect and the program
switch and control dial positions have not changed, the Sync Light will fl ash green
once when you re-attach the Neptune Connect to indicate that the sound processor
and Neptune Connect settings match and the parts are "synced." If the program
switch or control dial positions have been changed on the Neptune Connect before
re-attaching to the sound processor, the Sync Light will fl ash red once per second
until the settings have been synced.
The sound processor will not use any new program or control dial settings until it has
been synced with the Neptune Connect. To sync the Neptune Connect and sound
processor, fi rst turn the volume all the way down by moving the dial in a counter-
clockwise direction . The Sync Light will fl ash green once to indicate the Neptune
Connect and sound processor are synced and ready to apply setting changes. This
is an important safety feature preventing any sudden change in loudness or sound
quality if settings on the Neptune Connect have been altered prior to re-attaching
to the sound processor. After successfully syncing the sound processor and Neptune
Connect, return the volume dial to your preferred use setting.
Table 3: Neptune Connect Sync Light indicator
Behavior
One green fl ash
Flashes red once per second
Solid red light
Indication
Neptune Connect and processor settings are
synced. Any change will take effect on the
processor.
A mismatch exists between Neptune processor
settings and Neptune Connect settings. In
order to sync settings, turn volume dial all the
way down (counter-clockwise); verify green
light fl ashes once; then adjust the controls to
desired position(s).
Error condition. If this occurs, remove and
re-attach Neptune Connect.
User Guide for the Neptune™ Sound Processor 23