Technical Data
Speed range
Max. torque stirrer shaft
Permitted on - time
Speed adjustment
Speed display
Max. speed deviation (at nominal torque)
Design voltage
Design frequency
Input power (motor)
Output power (motor)
Max. output at stirrer shaft
Max. power input
Protection type to DIN EN 60 529
Excess voltage category
Protection class
Contamination level
Protection at overloaded
Perm ambient temperature
Perm. ambient humidity (rel.)
Operating position
Drive
Maximum stirring quantity- water
For medium viscosity (MV)
Clamping chuck- clamping range
Hollow shaft internal diameter Ø
Stand clamp
Housing
Max. dimension (W x D x H) without extension arm
Max. dimension (W x D x H) with extension arm
Weight with extension arm and clamping chuck
Operation at a terrestrial altitude
Subject to technical changes!
20
rpm
100 – 2000
Ncm
8
%
100
Rotary control knob (potentiometer) - front of appliance
Scale - front of appliance
%
± 5
VAC
220-240
100-120
Hz
50/60
50/60
W
28.5
28.5
W
17
17
W
17
17
W
45
45
IP 20
II
I (protective earth)
2
motor current limit
°C
+5 bis+40
%
80
on stand, chuck facing downwards
DC motor direct drive
ltr
8
mPas
1000-10000
mm
0.5 - 8
mm
no hollow shaft
bowl clamp
thermoplastic synthetic
mm
70 x 176 x 197
mm
70 x 336 x 197
kg
2
m
max. 2000 above sea level
032004