10. Calibration and repair of any instrument should only be performed by
qualified and trained service technicians.
11. Do not attempt calibration or service unless trained and another person
capable of rendering first aid and resuscitation is present.
12. Remember: Think Safety, Act Safely.
4) SAFETY INFORMATION:
The instrument complies with class II, overvoltage CAT IV 600V, CAT III
1000V of the IEC61010-1 (EN61010-1); IEC61010-2-032 (EN61010-2-032)
standards. Pollution degree 2 for Indoor Use. If the equipment is used in
a manner not specified, the protection provided by the equipment may be
impaired.
EMC: According with EN61326-1: 2006.
BY IEC61010 OVERVOLTAGE CATEGORY INSTALLATION
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equip-
ment supplied from the fixed installation.
NOTE: Examples include household, office and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed instal-
lation.
NOTE: Examples includes switches in fixed installations and some equip-
ment for industrial use with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at origin of the
installation.
NOTE: Examples include electricity meters and primary over-current pro-
tection equipment.
4