5.1
Electromagnetic Compatibility (EMC)
Portable and mobile RF communications equipment can affect the Affinity. Install and operate the Affinity according to the EMC
information presented in this chapter.
The Affinity has been tested for EMC emissions and immunity as a standalone instrument. Do not use the Affinity adjacent to or stacked
with other electronic equipment. If adjacent or stacked use is necessary, the user should verify normal operation in the configuration.
The use of accessories, transducers and cables other than those specified, with the exception of servicing parts sold by Interacoustics
as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the device.
Anyone connecting additional equipment is responsible for making sure the system complies with the IEC 60601-1-2 standard.
The Affinity is intended for use in the electromagnetic environment specified below. The customer or the user of the Affinity should assure that it is used in such an
environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Recommended separation distances between portable and mobile RF communications equipment and the Affinity.
The Affinity is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the Affinity can help prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Affinity as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum output power of
transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHZ, the higher frequency range applies.
Note 2 These guidelines may not apply to all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
The Affinity is intended for use in the electromagnetic environment specified below. The customer or the user of the
Affinity should assure that it is used in such an environment.
Immunity Test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and
voltage variations on power supply lines
IEC 61000-4-11
Power frequency (50/60 Hz)
IEC 61000-4-8
Note: UT is the A.C. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic emissions
Compliance
Group 1
Class B Limits
Complies
Class A Category
Complies
150 kHz to 80 MHz
d = 1.17
0.12
0.37
1.17
3.70
11.70
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
IEC 60601 Test
level
+6 kV contact
+8 kV air
+2 kV for power supply lines
+1 kV for input/output lines
+1 kV differential mode
+2 kV common mode
< 5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 sec
3 A/m
Electromagnetic environment - guidance
The Affinity uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The Affinity is suitable for use in all commercial, industrial, business, and residential
environments.
Separation distance according to frequency of transmitter
[m]
80 MHz to 800 MHz
d = 1.17
0.12
0.37
1.17
3.70
11.70
Compliance
+6 kV contact
+8 kV air
+2 kV for power supply lines
+1 kV for input/output lines
+1 kV differential mode
+2 kV common mode
< 5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT) for
5 cycles
70% UT (30% dip in UT) for
25 cycles
<5% UT
3 A/m
800 MHz to 2.5 GHz
d = 2.23
0.23
0.74
2.33
7.37
23.30
Electromagnetic
Environment-Guidance
Floors should be wood, concrete or
ceramic tile. If
floors are covered with synthetic material,
the relative humidity should be greater
than 30%.
Mains power quality should be that of a
typical commercial or residential
environment.
Mains power quality should be that of a
typical commercial or residential
environment.
(>95% dip in UT)
for 5 sec Mains power quality should be
that of a typical commercial or residential
environment. If the user of the Affinity
requires continued operation during
power mains interruptions, it is
recommended that the Affinity be
powered from an uninterruptable power
supply or its battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or residential environment.