MT10 Instruction for Use - EN
1.4
Warnings and Precautions
WARNING indicates a hazardous situation which, if not avoided, could result in
death or serious injury.
CAUTION, used with the safety alert symbol, indicates a hazardous situation
which, if not avoided, could result in minor or moderate injury.
NOTICE is used to address practices not related to personal injury
1. The instrument is not intended to be used in oxygen rich environments or use in conjunction
with flammable agents.
The instrument is not intended to be used in environments exposed to fluid spills.
1.
2. Ear tip are single use only.
3. Do not insert or in any way try to conduct measurements without proper probe eartip in place.
1. Be sure to insert the probe tip in a way which will assure an air tight fit without causing any
harm to the patient. Using a proper and clean ear tip is mandatory.
2. Be sure to use only stimulation intensities acceptable for the patient.
3. Never clean the transducer housing with water or insertion instruments.
4. Although the instrument fulfils the relevant EMC requirements precautions should be taken to
avoid unnecessary exposure to electromagnetic fields, e.g. from mobile phones etc. If the
device is used adjacent to other equipment it must be observed that no mutual disturbance
appears.
environment and consequently the health of human beings.
Within the European Union it is illegal to dispose electric and electronic
waste as unsorted municipal waste. Electric and electronic waste may
contain hazardous substances and therefore has to be collected separately.
Such products will be marked with the crossed-out wheeled bin shown
below. The cooperation of the user is important in order to ensure a high
level of reuse and recycling of electric and electronic waste. Failing to
recycle such waste products in an appropriate way may endanger the
Page 2