Therapy Life Indicators; Duration Of Prevena Plus Duo™ Therapy - KCI PREVENA PLUS DUO Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

THERAPY LIFE INDICATORS

Begin
Therapy
NOTE: Grey represents green indicators and black represents yellow indicators.
The therapy life indicators provide a visual display of the seven day therapy life cycle and the therapy
life remaining. When therapy begins all seven green LEDs are illuminated. During therapy, after each
24-hour period an indicator will turn off. When eight hours of therapy time remains, the last indicator
will illuminate with both a green and yellow LED simultaneously. When therapy time is about to
expire, the last indicator will illuminate with a yellow LED and an alert will sound for approximately
two minutes, then the therapy unit will shut off.
At the end of therapy, the therapy unit must be replaced with a new unit or alternative therapy must
be used. Patients should be instructed to contact the treating physician or caregiver if therapy unit
turns off and cannot be restarted before therapy is scheduled to end.
NOTE: Once therapy is on for one hour non-stop, the seven day lifespan begins and continues
even if the unit is turned off.
DURATION OF PREVENA PLUS DUO™ THERAPY
Therapy should be continuous for a minimum of two days up to a maximum of seven days.
NOTE: The PREVENA PLUS™ 125 Therapy Unit will automatically time-out after seven
days of therapy.
Patients should be instructed to contact their treating physician and not to turn therapy off
unless:
º
advised by the treating physician
º
bleeding develops suddenly or in large amounts during therapy
º
there are signs of allergic reaction or infection
º
the canister is full of fluid
º
system alerts must be addressed
Patient should be instructed to contact the treating physician if therapy unit turns off and
cannot be restarted before therapy is scheduled to end, or if canister becomes full of fluid.
At end of therapy, patient should return to treating physician for dressing removal.
4 Days
2 Days
Left
Left
1 Day
8 Hours
Left
16
Therapy ended.
Left
Therapy unit
will shut off.