5. Program
Program 1 - 10
Under menu "Programs", 10 user-defined pressure-rotation (rpm)-
time profiles can be created. In addition, the user can define whe-
ther the interval mode is activated or not in the programs.
If the interval mode is activated, the run time/stop time value from
"Interval Mode" setting will be taken.
Note: If the user needs to activate the interval mode in one seg-
ment of a program, he should set the "Ccw/cw time" in menu
option "Interval", meanwhile activate the "Ccw/cw" function in
menu option "Interval Mode" (see section "Rotation").
Select: Select the program.
Edit: Edit the selected program parameters. Start to edit the se-
lected program parameters by pressing menu option "Edit" with
start/stop knob. The user can edit, insert or delete one selected
program segment in the program. When user edits the program
time for at least one segment, a tick (√) for respective program
will appear.
Delete: Delete the selected program. If a selected program is dele-
ted by pressing on menu option "Delete" with start/stop knob, all
the program parameters will be emptied. The tick (√) disappears.
Sample for editing the program
22
Last measurement
Save as: Save the distillation sequence as a program.
Edit: Edit the selected program parameters.
Start to edit the selected program parameters by pressing on
menu option "Edit" with start/stop knob. The user can edit, insert
or delete one selected program segment in the program. When
the user edits the program time for at least one segment, a tick (√)
for the respective program will appear.
Delete: Delete the selected program. If a selected program is dele-
ted by pressing on menu option "Delete" with start/stop knob, all
the program parameters will be emptied. The tick (√) disappears.
Details for editing the program
When a program is being edited, the following screen appears.
In this program, user can define up to 10
segments. The selected segment is high-
lighted. Then, the user can edit, insert or
delete a segment in this program. When
the "Back" button is pressed after editing,
the program is saved automatically.
Edit: When the background of the selected value turns yellow the
user can change the pressure, the speed or the time.
Insert: With the insert option, a new segment will be insert below
the selected segment.
Delete: When deleting a highlighted segment, the settings will be
emptied. The section jumps to next segment.
Edit pressure
Edit speed
Edit time