1.2 Combination possibilities
This prosthetic component is part of the Ottobock modular system. As a res
ult of its design, it may be combined with other prosthetic components of the
modular system. You can find possible restrictions in this section. Contact
the manufacturer with questions.
2 Application
2.1 Indications for use
The product is to be used solely for lower limb prosthetic fittings.
2.2 Area of application
The product is recommended for mobility grade 1 (indoor walk
er). Approved for a body weight of max. 75 kg.
m°
kg
2.3 Environmental conditions
Allowable environmental conditions
Temperature range for use: -10 °C to +60°C
Allowable relative humidity 0 % to 90 %, non-condensing
Unallowable environmental conditions
Mechanical vibrations or impacts
Perspiration, urine, fresh water, salt water, acids
Dust, sand, highly hygroscopic particles (e.g. talcum)
Other environmental conditions
UV light
Hydrolysis
2.4 Service Life
This product has been tested by the manufacturer with 2 million load cycles.
Depending on the patient's activity level, this corresponds to a service life of
2 to 3 years.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or injury.
CAUTION
Warning regarding possible technical damage.
NOTICE
10 | Ottobock
Not resistant
Not resistant