Stereo microscope
Stemi 305
4.2.5
Controller K LED
The controller K (Fig. 24/4) is used to activate the
vertical illumination of the Stemi 305 body or the
ring illuminator in case of applications with boom
stands A or U or additional stands.
•
Mount the stands A and U including the
fastening of the Stemi 305 body to the Stemi
mount
as
described
instructions for the boom stands.
•
Screw-fasten front lens system 0.5x/185 mm
(Fig. 24/9) to the microscope body.
•
When using the ring illuminator, adapt the
latter to the front lens system (Fig. 24/9) using
the
adapter
supplied
(ID 58 mm, OD 66 mm).
•
Secure the adhesive side of the magnetic plate
(Fig. 24/7) to the swivel arm stand.
•
Plug the connection cable supplied with the
device (1 m long) (Fig. 24/2) into the socket
(Fig. 24/8) on the microscope body or the ring
illuminator and into the socket (Fig. 24/3) on
the back of the Controller K (Fig. 24/4).
•
Place Controller K (Fig. 24/4) next to the magnetic plate (Fig. 24/7).
•
Connect Controller K to the power supply using the power unit (Fig. 24/1).
•
Align swivel arm stand (Fig. 24/6) to enable an ergonomic working position.
•
Turn the knob (Fig. 24/5) on the Controller K to switch the vertical illuminator (or the ring illuminator)
ON and adjust illumination intensity by turning the knob.
•
Place specimen in the illuminated specimen field.
•
Set the stereo microscope to its initial setting, see section 4.1 on page 23.
04/2015
OPERATION
Setting reflected-light illumination
in
the
operating
with
the
device
435063-7044-010
Fig. 24
Controller K LED
ZEISS
27