If it is intended to operate the Stemi 305 with an external fibre optical cold light source, please
refer to the appropriate operating instructions incl. the safety provisions contained in them
before commissioning the light source.
The desktop power unit must not get into contact with moisture.
For transport over an extended distance, the device must be dismantled partially and
transported in its original packaging.
To transport the device over an extended distance, use the original packaging or the transport
case Stemi 305/508.
To connect the LED illuminators of the Stemi 305 system, six-pin modular connectors are used
which are known from telecommunications (type: RJ-12).
Only the LED illuminators pertaining to the Stemi 305 may be connected to the RJ-12 sockets in
the microscope body and in the stands of the Stemi 305. Only the connection cables supplied
with the device may be used.
The use of commercially available cables with RJ-12 connectors or the connection of fax units,
modems or other telecommunication equipment to the RJ-12 sockets of the Stemi 305 is
prohibited.
6
INTRODUCTION
Notes on instrument safety
435063-7044-010
Stereo microscope
Stemi 305
04/2015