Displaying/Selecting Insufflation Operating Mode; Setting The Nominal Pressure - General; Setting The Nominal Flow - General - Stryker PNEUMO SURE Instrucciones De Uso

Ocultar thumbs Ver también para PNEUMO SURE:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

5.6.2

Displaying/Selecting Insufflation Operating Mode

The device can be equipped with up to 4 procedure modes. Procedure options can
be released or uploaded in different combinations depending on the order or
software upgrade.
PNEUMO SURE
High Flow operating mode / Pediatric operating mode
PNEUMO SURE XL
High Flow operating mode / Pediatric operating mode /
Bariatric operating mode / Vessel Harvest operating
mode
1. Control or selection of the insufflation modes can be done only while insuf-
flation is stopped. Press the function field for 2 s to change the insufflation
operating mode.
2. The display depicts an insufflation operating modes overview. Unavailable
operating modes are depicted in gray and cannot be selected. Press the re-
spective function field to select the corresponding desired operating mode
(e.g. Pediatric).
3. After pressing the function field for the insufflation operating mode, the dis-
play depicts the selected procedure (e.g. Pediatric operating mode). The dis-
played parameters correspond with the factory settings or the values set in
the Configuration menu (see chapter 10.1 Configuration menu I, page 60).
5.6.3

Setting the Nominal Pressure - General

Press the function fields º or » on the display below the text line Set Pressure to
set the nominal pressure.
Applies to all modes:
• The nominal pressure can be increased or decreased during insufflation or
while insufflation is stopped.
• With every touch of the pressure function field º or » the nominal pressure is
increased/reduced in steps of 1 mm Hg. When pressing the pressure function
field º or » longer than 1,5 seconds, scrolling is activated but only up to the re-
spective, application-dependent safety threshold. Once this limit has been
reached, the status line depicts the message Safety limit and the actual value
flashes. To allow settings above the safety threshold release the pressure func-
tion field for 2 seconds. The display switches back to the status display. Now it
is possible to increase the pressure above the safety threshold but only to the
next safety threshold (as far as existent) or to the maximum setting.
• The latest working settings are stored with interruption of the insufflation, but
only if the last value was below the lowest safety threshold (15 mm Hg for High
Flow and Bariatric operating mode, 12 mm Hg with Pediatric operating mode
and Vessel Harvest operating mode). Otherwise the set pressure value is reset
to the lowest safety threshold upon deactivation of insufflation.
• The latest settings are reset to the values preselected in the Configuration
menu when deactivating the device or when changing into the Configuration
menu or the Mode Selection menu. Upon choice of the respective working
mode the nominal settings as stored in the Configuration menu are displayed
(see chapter 10.1 Configuration menu I, page 60).
5.6.4

Setting the Nominal Flow - General

Press the function fields º or » on the display below the text line Set Flow to set
the nominal flow.
Applies to all modes:
Operating the Device - General
Set Pressure
Actual Pressure
Set Flow
mm Hg
mm Hg
l/min
0
15
3
0.0
Liter
Mode: High Flow
(1)
Change
Actual
START
0
Mode
l/min
(2)
EN
23