12.4

Replacing the Fuse

CAUTION!
Before replacing the fuse, check the values of the fuse to be inserted acc. to chap-
ter 16 Technical Data, page 89.
The fuse may be defective and is in need of replacement if:
• displays and LEDs do not light up,
• the device does not function.
Check to make sure
• the main power supply cable is properly connected to the power supply input
and to a safety socket,
• the house power supply fuse is functioning.
WARNING!
Unplug the power cable from the device before checking the fuse.
The device does not have to be opened to replace the fuse.
1. Switch device off.
2. Disconnect device from power supply.
3. Remove power connection cable from mains socket.
4. The fuse holder is located to the left of the mains socket. Remove fuse holder
as depicted in Fig. 12-1 .
C
5. A Undo the latch of the fuse holder with a small screwdriver.
6. B Remove the fuse holder.
7. C Check fuses.
8. Insert a new fuse. Use only the specified type of fuse (see chapter 16 Techni-
cal Data, page 89).
9. Insert the fuse holder until it can be heard snapping into place.
10. Use the power cable to reconnect the shockproof safety socket with the rear
mains socket.
12.5

Care and Maintenance of Reusable Tube Set

The device can be used with reusable tube set with gas heating validated for this
purpose. Please observe the following notices when using a reusable tube set.
A
B
Care and Maintenance
Fig. 12-1 Opening the fuse holder
EN
75