Cleaning
8
YCP06SQ Operating Instructions

5.2 Pipette Calibration

Prerequisites
Test liquid: Use only de-aerated, bi-distilled water at
room temperature.
Procedure
Calibration of the pipette: The max. filling quantity
t
of 50 ml must not be exceeded.
If necessary, empty the pipetting vessel.
t
5.3 Adjustment and Calibration of the
Balance
Procedure
Switch off the "isoCAL" function before pipette
t
calibration (see operating instructions for the
balance).
Before starting the adjustment/calibration: Reinstall
t
any uninstalled parts in the weighing chamber of
the balance along with the weighing pan.
6
Cleaning
Notice
Do not clean the pipette calibration kit in the
dishwasher.
Procedure
Clean the parts with a mild cleaning agent
t
(e.g. isopropanol)
After cleaning, dry the parts with a lint-free cloth.
t
7