Preventative Maintenance Checklist

All fasteners secure
All welds intact, not cracked or broken
No bent or broken tubing or sheet metal
No debris in casters
All casters secure and swivel properly
Engage brake pedal and push on the bed to ensure all casters lock securely
Steer caster latches properly
Siderails move and latch properly
Fowler operates properly
Fowler (back rest) Slide operates properly
Bed Up/Down operates properly
Foot section operates properly
Foot uprights operate properly
Trendelenburg operates properly
IV pole intact and operating properly
No rips or cracks in mattress cover
Lubricate where required
Replace Nurse Call 9V battery (annually)
Power cord not frayed
No cables worn or pinched
All electrical connections tight
All grounds secure to the frame
Ground impedance not more than 100 milliohms
Current leakage not more than 300 microamps
Ensure ground chains are clean, intact, and have at least two links touching the floor.
Serial No.
NOTE
Preventative maintenance should be performed at a minimum of annually. A preventative maintenance pro-
gram should be established for all Stryker Medical equipment. Preventative maintenance may need to be
performed more frequently based on the usage level of the product.