10. Troubleshooting

Problem
If the motor is not
running
If there is low
or no suction
If condensation
appears in
the tubing
The pump
becomes wet
Solution
Check presence of mains current.
Check the position of the batteries in the battery compartment.
Verify the membrane is clean and not damaged.
Make sure that the membrane forms a seal around the edge of the
connector and the lid is firmly connected with the connector.
Check all connections.
Check if the ends of the tubing fit securely into the back of the connector
and into the port on the breast pump. During single pumping, insert the
unused tubing into the tubing holder.
There should be no water in the tubing.
Do not kink the tubing during expression.
Ensure the breast shield forms a complete seal around the breast.
All components must be completely dry.
Clean the tubing as described in chapter 6.5. Shake out any water drops.
Unplug the mains adapter from the socket.
Switch off the pump.
Do not place the pump upside down. The control panel must point upwards.
Store the pump in a dry, warm place for 24 hours.
Note:
Do not expose the pump to direct sunlight.
Do not accelerate the drying process by use of heat.