Operation

6
The meter is a microprocessor controlled device for the measurement of
pH value, redox potential and temperature, as well as the simulation of pH value
and redox.
6.1
Press the key "pH/mV/°C" to select the measure pH operating state.
LCD marker "pH" is activated.
The measured pH value is temperature-compensated. If a Pt 1000 probe is
connected the measured temperature value is used for temperature-
compensation. Otherwise the temperature can be set manually.
If measured value is out of range LCD markers "ERR" and "pH" will flash.
6.2
Connect the probe to the "meas" connector.
Press the key "pH/mV/°C" to select the measure redox operating state.
LCD marker "mV" is activated.
The meter displays the redox potential (no user calibration is possible).
If measured value is out of range LCD markers "ERR" and "mV" will flash.
6.3
Only Pt 1000 probes can be connected for temperature measurement. Not
suitable for thermocouples or Pt 100 probes.
Connect the Pt 1000 probe to the "Sim/Temp" connector.
Press the key "pH/mV/°C" to select the "measure Pt 1000" operating state.
LCD marker "°C" is activated.
The meter displays the temperature in °C (no user calibration is required).
If measured value is out of range LCD markers "ERR" and "°C" will flash.
34
Operation
NOTE
Please read through the function description section first in order
to learn how to use the meter and change values and operating
modes.

pH measurement

Redox measurement

Temperature measurement

®