Ph Calibration - ProMinent DULCOMETER Pt 1000 Instrucciones De Servicio

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6.4
6.4.1
Calibrating the meter:
Immerse the pH probe into the buffer solution and stir gently.
Press the "CAL" key: the LCD elements "cal" and "pH" are steadily
displayed and element "P1" (i.e. buffer 1) is flashing.
The buffer value 7.00 pH is indicated on the LCD display.
Press the CAL key repeatedly to move through buffer values 4.00/7.00/
10.00 pH.
Press the keys "UP" or "DOWN" to change the selected buffer value in
0.01 pH steps, if required.
If the calibration is already in operation when the key is pressed, it will be
started anew; thus the calibrating buffer and its exact value can be
changed after start of the actual calibration.
Actual calibration starts 2 sec after pressing the last key; the LCD marker
"P1" is displayed steadily.
After 30 sec. the displayed buffer value flashes if the calibration of buffer 1
is still in operation.
If the calibration of buffer1 is not valid, an error message is displayed. See
Error messages when calibrating.
If the calibration of buffer 1 is valid the LCD element "P2" is switched on
and flashes in addition to element "P1".
Select a new buffer using the key CAL to start the calibration of buffer 2.
®
ProMinent

pH calibration

Two point pH calibration
CAUTION
• The pH sensors must be calibrated at regular intervals, i.e.
initially 24 hours after the original calibration and thereafter
each week.
• Always use high quality pH buffer solutions to calibrate the
pH sensors! The "Phenol red method" is too imprecise for
this calibration!
Operation
35

Publicidad

Tabla de contenido
loading

Tabla de contenido